Hurricane Bone Collector [KAB001]
An instrument that thoroughly retrieves the autogenous bone fragment
and adult stem cells that have been discarded through the suction
during drilling of the alveolar bone for implant placement.
Composition
Assembly Method
- ① Centrifugal Tube & Cup assembly in Hurricane Body
- ② Assemble Suction Handle & Tip in use to Hurricane Bone Collector
Weight
- Hurricane bone collector(Body+Tube+Cup) : 117g
- Hurricane bone collector + Suction(Handle & Tip) : 200g
Specimen Weight
- Centrifugal Tube Specimen Volume :12ml
Operating Principle
When coolant, blood, fat, bone fragments etc. are sucked duringdrilling
inside the Hurricane Bone Collector body, the descending vortexformed by centrifugal force
gathers bone fragment, bone dust, blood, stem cells in Centrifugal Cup.
Instruction
General suction instrument
is used for general medical
trea tments such as incision
valve formation and
inflammation removal.Hurricane Suction is mount-
ed just before drilling for the
placement of Implant.Adjust the suction tip down-
wards to fit mandible.Position the specimen colle-
ction tube downwards.Bone tissue is collected dur-
ing drilling with sufficient
saline solution douche.
- Insufficient saline solution
cause blood coagulation
and disturbs collecting
bony tissue.Disassemble the Hurricane
Suction and collection tube
after completion of drillingReassemble the Suction Tip
to original Suction
instrument.Place two specimen tube in
symmetrical position and do
centrifugation under the same
conditions as PRF.Remove supernatant inside
the collection tube through
suction.Take out the bone tissue and
PRF from the bottom of the
collection tube.Loss of Implant buccal plate
bone.Bone graft with INNO CaP.
Graft Hurricane autogenous
bone tissue.Graft autogenous bone tissue
on outermost layer.
Way to Grasp
-
Example of correct way to grasp
Comfortably grip in an angle which assistant do not interfere with
the operator and patient but maintain the Centrifugal Tube
in vertical direction. -
Example of incorrect way to grasp
Depending on the situation, the instrument may work correctly
at an incline of less than 15 ° in the vertical direction,
but in larger angles, the instrument may not work properly.
Suction Tip Direction Adjustment Method
Following to the position of the operator(doctor) and the assistant(hygienist), rotate the direction of the Suction Tip
according to the patient's intraoral surgical site.
Centrifuge Application Method
Use after purifying the collected specimen from the centrifuge.
① Insert the Centrifugal Cup
downwards.
② Symmetrically, both horizontally and vertically,
insert the CentrifugalTube in the Centrifuge.
-
Caution A
- The maximum volume of collected specimens is the amount of
specimen generated during the operation of 2 or less implant
placementper Centrifugal tube.
- The maximum specimen volume to be collected in Centrifugal
tube is 12ml, however appropriate volume of specimen is
8~10ml of specimen to prevent flowing down during movement
or operating centrifuge.
- The maximum volume of collected specimens is the amount of
-
Caution B
- In case of different height of collected specimen in centrifugal Tube,
add saline solution to adjust the level. Different height may cause
unsatisfactory purification effect due to not properly
functioning centrifuge.
- In case of different height of collected specimen in centrifugal Tube,
Cleaning and Sterilization Methods for Reuse
< Cleaning >
- 1. After the procedure, suck 1L of distilled water or saline solution with suction device
to remove blood and foreign substances inside the instrument. - 2. After disassembling the Hurricane adapted body, tube and cup,
following to the instruction of manufacturer, dilute the enzyme
cleaner with tap water . Then proceed ultrasonic
cleaning for 10 minutes then clean with tap water for 3 minutes. - 3. Completely drain with dry cloth or fan heater.
< Sterilization >
- 1. Before use, sterilize for 15 minutes at 132℃ or at 135℃
for 10 minutes with a high pressure steam sterilizer . - 2. Dry for 30 minutes after high pressure steam sterilizer.
- 3. Keep the sterilized instruments in clean environment at 15-30℃.